** The EnviroGard, 2,4-D Test kit used in data analysis
from Millipore is based on the use of polyclonal antibodies which bind
either 2,4,-D residues or 2,4-D-enzyme conjugate. The 2,4-D from
the sample competes with 2,4-D-enzyme conjugate for a limited number of
antibody binding sites. Antibodies that bind 2,4-D are immobilized
to the walls of the test tubes.
Since there are the same number of
antibody binding sites on every test tube, and each test tube receives
the same number of 2,4-D-enzyme conjugate molecules, a sample which contains
a low concentration of 2,4-D enables the antibody to bind many 2,4-D-enzyme
conjugate molecules. Therefore, a low concentration of 2,4-D will
allow fewer 2,4-D-enzyme conjugate molecules to be bound by the antibodies.
Tube Marking Tube Contents
"-"
negative control
"+"
2.0 or 50 ppb calibrator
S1
sample 1
S2
sample 2
S3
sample 3
S4
sample 4
S5
sample 5
2. Add 4 drops (160
ul) of sample or negative control or 50 ppb calibrator depending on test
tube
3. Add 4 drops (160 ul) of enzyme
conjugate to each test tube.
4. Gently mix and let stand
for 10 minutes.
5. Fill the test tubes to overflowing
with tap or distilled water, decant and vigorously shake out the
water.
6. Repeat rinse 4 times.
7. Add 4 drops of substrate
to each test tube.
8. Immediately add 4 drops of
chromogen to test tubes.
9. Mix by shaking and let sit
for 10 minutes.
10. Add 1 drop of STOP solution
to each test tube. - Solution will turn yellow.
11. Add 500 ul of distilled
water prior to reading on the Spectrophotometer 20.
12. Measure and record the absorbence
at 450nm of the negative control, each sample, and the calibrator.
Calculations:
Compare the OD of each sample to the OD of the calibrator. If the OD of the sample is less than the OD of the calibrator, the sample contains a concentration greater than the concentration of that calibrator. Conversely, if the OD of the sample is greater than the OD of a calibrator, the sample contains a concentration that is lower than the concentration found in the calibrator.
Example: OD Interpretation
Neg. Ctrl.
1.298
2.0 ppb
1.068
50 ppb
0.439
Sample
0.755
>2.0 ppb; <50 ppb